Achieving safe hands-on defibrillation using electrical safety gloves – A clinical evaluation

RESUSCITATION JOURNAL COVER

Achieving safe hands-on defibrillation using electrical safety gloves – A clinical evaluation

Charles D. Deakin, Jakob E. Thomsen, Bo Løfgren, Graham W. Petley

 

Abstract

 

Introduction

Safe hands-on defibrillation (HOD) will allow uninterrupted chest compression during defibrillation and may improve resuscitation success. We tested the ability of electrical insulating gloves to protect the rescuer during HOD using a ‘worst case’ electrical scenario.

Materials and method

Leakage current flowing from the patient to the ‘rescuer’ during antero-lateral defibrillation of patients undergoing elective cardioversion was measured. The ‘rescuer’ maintained firm (20?kgf) contact with the patient during defibrillation, wearing Class 1 electrical insulating gloves while simulating an inadvertent contact with the patient, through an additional wired contact between ‘rescuer’ and patient.

Results

Data from 61 shocks from 43 different patients were recorded. The median leakage current from all defibrillations was 20.0??A, (range: 2.0–38.5). In total, 18 of the shocks were delivered at 360?J and had a median leakage current of 27.0??A (range: 14.3–38.5).

Conclusion

When using Class 1 electrical insulating gloves for hands-on defibrillation, rescuer leakage current is significantly below the 1?mA safe threshold, allowing safe hands-on defibrillation if the rescuer makes only one other point of contact with the patient.